Technical Specifications

oCelloScope

Measurement Specifications
Detectable object size0.5 µm – 2 mm
Optical Magnification4x
Optical resolution1.3 µm
Optical PrincipleFluidScope TM
Size, power specifications
Dimensions (L x W x H)450 x 260 x 250 mm (with open lid: H = 550 mm)
Weight9.6 kg
Power adapter, inputInput : 100-240 V, ~50/60 Hz, 600-300 mA
Output : 24 V, 1250 mA

Can be ordered with EU, US/JP, UK, or China outlet plugs
or with IEC connector for use with an IEC C7 power cables
Power useIdle state : 11 W
Active state : 20 W
ConnectorsRJ45 (1 Gigabit Ethernet)
Environmental Specifications (surrounding conditions, does not include the sample to be measured)
Operating temperature5-40 °C
Operating humidity20-93% RH

ParticleTech Sample Unit

Size, power specifications
Dimensions (L x W x H)400 x 250 x 165 mm
Weight8.9 kg
Power adapterInput : 100-240 V, ~ 50/60 Hz, 1.5 A Max
Output : 12 V, 5 A

The power supply supports C13 power cables
Can be ordered with EU, US/JP, and UK cables
ConnectorsUSB 2.0

Flow Cell

Size, material specifications
Dimensions (L x W x H)76 x 41 x 4.5 mm
Chamber heights 1100, 800, 100 µm
Material, foil (chamber)PMMA

Flow Cell is generally not affected by:
Alkalis
Detergents
Dilute inorganic acids
Aliphatic hydrocarbons
Dilute alcohols

Flow Cell is generally not recommended for use with:
Esters
Ketones
Aromatic hydrocarbons
Chlorinated hydrocarbons

Temperature: up to 80°C
TemperatureUp to 80°C
Tubes to connect to Flow Cell (also ideal for pumps)Flexible transparent silicone tubing
Inner diameter3.2 mm
Wall thickness1.6 mm

Flow Cell Adapter

Size specifications
Dimensions (L x W x H)127 x 85 x 14 mm
Weight0.16 kg